8: TSP command reference Model 2461 Interactive SourceMeter® Instrument
8-10 2461-901-01 A/November 2015
Example
Autocalibration starts running.
Also see
Autocalibration (on page 3-51)
beeper.beep()
This function generates an audible tone.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
beeper.beep(duration, frequency)
The amount of time to play the tone (0.001 to 100 s)
The frequency of the beep (20 to 8000 Hz)
Details
You can use the beeper of the instrument to provide an audible signal at a specific frequency and
time duration. For example, you can use the beeper to signal the end of a lengthy sweep.
Using this function from a remote interface does not affect audible errors or key click settings that
were made from the Model 2461 front panel.
Example
Generates a 2 s, 2400 Hz tone.
Also see
None
buffer.clearstats()
This function clears the statistical information associated with the specified buffer.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
buffer.clearstats()
buffer.clearstats(bufferVar)
The name of the reading buffer, which may be a default buffer (defbuffer1 or
defbuffer2) or a user-defined buffer; defaults to defbuffer1 if not specified
Details
This command clears the statistics without clearing the readings.