8: TSP command reference Model 2461 Interactive SourceMeter® Instrument
8-36 2461-901-01 A/November 2015
bufferVar.statuses
This attribute contains the status values of readings in the reading buffer.
Type TSP-Link accessible Affected by Where saved Default value
Instrument reset
Power cycle
Usage
statusInformation = bufferVar.statuses[N]
The status value when reading N of the specified buffer was acquired; refer to
The name of the reading buffer, which may be a default buffer (defbuffer1 or
defbuffer2) or a user-defined buffer
The reading number N; can be any value from 1 to the number of readings in the
buffer; use the bufferVar.n command to determine the number of readings in
Details
This command is not available if the buffer style is set to compact.
This buffer recall attribute holds an array (a Lua table) of the status values for all the readings in the
buffer. The status values are floating-point numbers that encode the status value. Refer to the
following table for values.
Buffer status bits for sense measurements
Bit (hex) Name Decimal Description
statusInformation parameter
1 Measure status
questionable
6 A/D converter
from which
reading
originated; for
the Model 2461,
this will always
be 0 (main) or 2
terminal, front is
Measure status
limit 2 low
32 Measure status
Measure status
limit 1 low
Measure status
limit 1 high