8: TSP command reference Model 2461 Interactive SourceMeter® Instrument
8-320 2461-901-01 A/November 2015
tsplink.line[N].reset()
This function resets some of the TSP-Link trigger attributes to their factory defaults.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
tsplink.line[N].reset()
The trigger line (1 to 3)
Details
The tsplink.line[N].reset() function resets the following attributes to their default values:
• trigger.tsplinkin[N].edge
• trigger.tsplinkout[N].logic
• tsplink.line[N].mode
• trigger.tsplinkout[N].stimulus
• trigger.tsplinkout[N].pulsewidth
This also clears trigger.tsplinkin[N].overrun.
Example
Resets TSP-Link trigger line 3
attributes to default values.
Also see
trigger.tsplinkin[N].edge (on page 8-311)
trigger.tsplinkin[N].overrun (on page 8-311)
trigger.tsplinkout[N].logic (on page 8-313)
trigger.tsplinkout[N].pulsewidth (on page 8-314)
trigger.tsplinkout[N].stimulus (on page 8-315)
tsplink.line[N].mode (on page 8-319)
tsplink.line[N].state
This attribute reads or writes the digital state of a TSP-Link synchronization line.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Not applicable Not applicable tsplink.STATE_HIGH
Usage
lineState = tsplink.line[N].state
tsplink.line[N].state = lineState
The state of the synchronization line:
• Low: tsplink.STATE_LOW
•
The trigger line (1 to 3)