Model 2461
Interactive SourceMeter® Instrument Reference Manual Section 8:
TSP command reference
2461-901-01 A/November 2015 8-21
bufferVar.endindex
This attribute indicates the last index in a reading buffer.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R)
Yes
Restore configuration
Instrument reset
Power cycle
Not applicable
Not applicable
Usage
bufferVar.endindex = endIndex
endIndex
Ending index of the buffer
bufferVar
The name of the reading buffer, which may be a default buffer (defbuffer1 or
defbuffer2) or a user-defined buffer
Details
Use this attribute to find the ending index in a reading buffer.
Example
test1 = buffer.make(100)
smu.measure.count = 6
smu.measure.read(test1)
print(test1.startindex, test1.endindex, test1.capacity)
smu.measure.read(test1)
print(test1.startindex, test1.endindex)
Create a buffer named test1 with a capacity of 100 readings.
Set the measure count to 6.
Make measurements and store them in buffer test1.
Get the start index, end index, and capacity of test1.
Output: 1, 6, 100
Make six more measurements and store them in buffer test1.
Get the start index and end index of test1.
Output: 1, 12
Also see
bufferVar.startindex (on page 8-35)
buffer.make() (on page 8-14)
Reading buffers (on page 3-2)
Remote buffer operation (on page 3-23)

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
BrandKeithley
Model2461
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals