Interactive SourceMeter® Instrument Reference Manual Section 6:
2461-901-01 A/November 2015 6-121
:SOURce[1]:PULSe:TRain:<function>
This command defines a sequence of source pulses and creates a trigger model to generate the pulse train.
Type Affected by Where saved Default value
Command only
Recall settings
Instrument reset
Save settings
Source configuration list
Measure configuration list
Not applicable
Usage
:SOURce[1]:PULSe:TRain:<function> <biasLevel>, <pulseLevel>, <pulseWidth>, <count>
:SOURce[1]:PULSe:TRain:<function> <biasLevel>, <pulseLevel>, <pulseWidth>, <count>,
<measEnable>
:SOURce[1]:PULSe:TRain:<function> <biasLevel>, <pulseLevel>, <pulseWidth>, <count>,
<measEnable>, "<bufferName>"
:SOURce[1]:PULSe:TRain:<function> <biasLevel>, <pulseLevel>, <pulseWidth>, <count>,
<measEnable>, "<bufferName>", <delay>
:SOURce[1]:PULSe:TRain:<function> <biasLevel>, <pulseLevel>, <pulseWidth>, <count>,
<measEnable>, "<bufferName>", <delay>, <offTime>
:SOURce[1]:PULSe:TRain:<function> <biasLevel>, <pulseLevel>, <pulseWidth>, <count>,
<measEnable>, "<bufferName>", <delay>, <offTime>, <xBiasLimit>
:SOURce[1]:PULSe:TRain:<function> <biasLevel>, <pulseLevel>, <pulseWidth>, <count>,
<measEnable>, "<bufferName>", <delay>, <offTime>, <xBiasLimit>, <xPulseLimit>
:SOURce[1]:PULSe:TRain:<function> <biasLevel>, <pulseLevel>, <pulseWidth>, <count>,
<measEnable>, "<bufferName>", <delay>, <offTime>, <xBiasLimit>, <xPulseLimit>,
<failAbort>
The source function:
• Voltage pulse sweep: VOLTage
•
Current pulse sweep: CURRent
Output level the instrument sources before the first pulse and returns to between
pulses (delay and off time):
• Current: -7.35 A to 7.35 A
•
The amplitude current or voltage from zero (not from the bias level):
• Current: -10.5 A to 10.5 A
•
The time at the amplitude level for each pulse:
• Extended operating area: 10 A at 100 V, 5 % duty cycle, maximum pulse
width: 1 ms, minimum pulse width 150 µs (load dependent)
• Normal operating area (DC): 99.99 % duty cycle, maximum pulse width
10,000 s, minimum pulse width 150 µs
The number of pulses in the pulse train; default is 1:
• Infinite loop: 0
• Finite loop:
to
Enable or disable measurements at the top of each pulse:
• Enable: ON (default)
• Disable:
A string that indicates the reading buffer; the default buffers (defbuffer1 or
defbuffer2) or the name of a user-defined buffer; if no buffer is specified,
is used
The amount of time that the instrument stays at bias level before each pulse (in
seconds): 0 to 10,000 s