8: TSP command reference Model 2461 Interactive SourceMeter® Instrument Refer
8-238 2461-901-01 A/November 2015
trigger.blender[N].reset()
This function resets some of the trigger blender settings to their factory defaults.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
trigger.blender[N].reset()
The trigger event blender (1 or 2)
Details
The trigger.blender[N].reset() function resets the following attributes to their factory
defaults:
• trigger.blender[N].orenable
• trigger.blender[N].stimulus[M]
It also clears trigger.blender[N].overrun.
Example
trigger.blender[1].reset()
Resets the trigger blender 1
settings to factory defaults.
Also see
trigger.blender[N].orenable (on page 8-236)
trigger.blender[N].overrun (on page 8-237)
trigger.blender[N].stimulus[M] (on page 8-238)
trigger.blender[N].stimulus[M]
This attribute specifies the events that trigger the blender.
Type TSP-Link accessible Affected by Where saved Default value
Instrument reset
Power cycle
Trigger blender N reset
Usage
event = trigger.blender[N].stimulus[M]
trigger.blender[N].stimulus[M] = event
The event that triggers the blender action; see Details
An integer that represents the trigger event blender (1 or 2)
An integer representing the stimulus index (1 to 4)
Details
There are four stimulus inputs that can each select a different event.
Use zero to disable the blender input.