Interactive SourceMeter® Instrument Reference Manual Section 8:
2461-901-01 A/November 2015 8-175
smu.reset()
This function turns off the output and resets the commands that begin with smu. to their default settings.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
smu.reset()
Details
This function turns off the output and resets the commands that begin with smu. to their default
settings.
Example
Turns off the output and resets the SMU
commands to their default settings.
Also see
reset() (on page 8-97)
smu.source.autorange
This attribute determines if the range is selected manually or automatically for the selected source function or
voltage source.
Type TSP-Link accessible Affected by Where saved Default value
Instrument reset
Power cycle
Source configuration list
Source configuration list
Usage
sourceAutorange = smu.source.autorange
smu.source.autorange = sourceAutorange
Disable automatic source range: smu.OFF
Enables automatic source range:
Details
This command indicates the state of the range for the selected source. When automatic source range
is disabled, the source range is set manually.
When automatic source range is enabled, the instrument selects the range that is most appropriate
for the value that is being sourced. The output level controls the range. If you read the range after the
output level is set, the instrument returns the range that the instrument chose as appropriate for that
source level.
If the source range is set to a specific value from the front panel or a remote command, the setting for
automatic range is set to disabled.
Only available for the current and voltage functions.