8: TSP command reference Model 2461 Interactive SourceMeter® Instrument
Reference Manual
8-214 2461-901-01 A/November 2015
smu.source.sweeplist()
This function sets up a sweep based on a configuration list, which allows you to customize the sweep.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
smu.source.sweeplist(configListName)
smu.source.sweeplist(configListName, index)
smu.source.sweeplist(configListName, index, sDelay)
smu.source.sweeplist(configListName, index, sDelay, count)
smu.source.sweeplist(configListName, index, sDelay, count, failAbort)
smu.source.sweeplist(configListName, index, sDelay, count, failAbort, bufferName)
configListName
The name of the source configuration list that the sweep uses; this must be defined
before sending this command
index
The index in the configuration list where the sweep starts; default is 1
sDelay
The delay between measurement points; default is 0 for no delay or you can set a
specific delay value from 50 µs to 10 ks
count
The number of times to run the sweep; default is 1:
Infinite loop: smu.INFINITE
Finite loop:
1
to
268,435,455
failAbort
Complete the sweep if the source limit is exceeded: smu.OFF
Abort the sweep if the source limit is exceeded:
smu.ON
(default)
bufferName
Indicates the reading buffer to use; the default buffers (defbuffer1 or
defbuffer2) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1 is used
Details
This command allows you to set up a custom sweep, using a configuration list to specify the source
levels.
When you specify a delay, a delay block is added to the sweep trigger model. This delay is added to
any source delay you may have set. For example, if you set 10 ms for the source delay and 25 ms for
the delay in the for the log sweep command, the actual delay is 35 ms.
To run the sweep, initiate the trigger model.

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
BrandKeithley
Model2461
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals