Model 2461
Interactive SourceMeter® Instrument Reference Manual Section 8:
TSP command reference
2461-901-01 A/November 2015 8-207
Reset the instrument to its default settings.
Set the source function to current and turn on source readback.
Set the digitize function to voltage and the digitize voltage range to 20 V.
Set the digitizer sample rate to 25000 readings per second.
Set up a pulse train with a bias level of -1 A and a pulse level of -3 A. Set the train to create three pulses with
a pulse length of 10 ms and an off time of 90 ms. Set the pulse voltage limit and the bias voltage limit to 15 V.
Start the pulse train.
Also see
Operating boundaries (on page 4-4)
Pulse operation (on page 3-64)
smu.source.pulsesweeplist() (on page 8-200)
Trigger model (on page 3-107)
smu.source.range
This attribute selects the range for the source for the selected source function.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Source configuration list
Function change
Configuration script
Source configuration list
Current: 1 µA
Voltage: 200 mV
Usage
rangeValue = smu.source.range
smu.source.range = rangeValue
rangeValue
Set to the maximum expected voltage or current to be sourced; see Details for
values; the ranges are:
Current: -10 A to 10 A
Voltage: -100 V to 100 V
Details
This command manually selects the measurement range for the specified source.
If you select a specific source range, the range must be large enough to source the value. If not, an
overrange condition can occur.
If an overrange condition occurs, an event is displayed and the change to the setting is ignored.
The fixed current source ranges are 1 µA, 10 µA, 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 4 A, 5 A, 7 A,
and 10 A. Note that if the 10 A range is selected, the maximum source output level is 7.35 A unless
pulsing is active.
The fixed voltage source ranges are 200 mV, 2 V, 7 V, 10 V, 20 V, and 100 V.
When you read this value, the instrument returns the positive full-scale value that the instrument is
presently using.

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
BrandKeithley
Model2461
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals