Interactive SourceMeter® Instrument Reference Manual Section 6:
2461-901-01 A/November 2015 6-61
Example 1
Sets the measurement time for current
measurements to 0.0083 s (0.5/60).
Example 2
Sets the measurement time for resistance
measurements to 0.0083 s (0.5/60).
Example 3
Sets the measurement time for voltage
measurements to 0.0083 s (0.5/60).
Also see
Using NPLCs to adjust speed and accuracy (on page 4-9)
[:SENSe[1]]:<function>:OCOMpensated
This command enables or disables offset compensation.
Type Affected by Where saved Default value
Instrument reset
Power cycle
Measure configuration list
Measure configuration list
Usage
[:SENSe[1]]:<function>:OCOMpensated <state>
[:SENSe[1]]:<function>:OCOMpensated?
• Current: CURRent[:DC]
• Resistance: RESistance
•
Disable offset compensation: OFF or 0
Enable offset compensation: ON or 1
Details
The voltage offsets caused by the presence of thermoelectric EMFs (V
EMF
) can adversely affect
resistance measurement accuracy. To overcome these offset voltages, you can use offset-
compensated ohms.
This feature is only applied to resistance measurements or when [:SENSe[1]]:CURRent:UNIT or
[:SENSe[1]]:VOLTage:UNIT is set to OHM.