Model 2461
Interactive SourceMeter® Instrument Reference Manual Section 6:
SCPI command reference
2461-901-01 A/November 2015 6-61
Example 1
CURR:NPLC 0.5
Sets the measurement time for current
measurements to 0.0083 s (0.5/60).
Example 2
RES:NPLC 0.5
Sets the measurement time for resistance
measurements to 0.0083 s (0.5/60).
Example 3
VOLT:NPLC 0.5
Sets the measurement time for voltage
measurements to 0.0083 s (0.5/60).
Also see
Using NPLCs to adjust speed and accuracy (on page 4-9)
[:SENSe[1]]:<function>:OCOMpensated
This command enables or disables offset compensation.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
OFF (0)
Usage
[:SENSe[1]]:<function>:OCOMpensated <state>
[:SENSe[1]]:<function>:OCOMpensated?
<function>
The measure function:
Current: CURRent[:DC]
Resistance: RESistance
Voltage: VOLTage[:DC]
<state>
Disable offset compensation: OFF or 0
Enable offset compensation: ON or 1
Details
The voltage offsets caused by the presence of thermoelectric EMFs (V
EMF
) can adversely affect
resistance measurement accuracy. To overcome these offset voltages, you can use offset-
compensated ohms.
This feature is only applied to resistance measurements or when [:SENSe[1]]:CURRent:UNIT or
[:SENSe[1]]:VOLTage:UNIT is set to OHM.

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
BrandKeithley
Model2461
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals