8: TSP command reference Model 2461 Interactive SourceMeter® Instrument
8-32 2461-901-01 A/November 2015
Example
testData = buffer.make(50)
smu.source.output = smu.ON
trigger.model.load("SimpleLoop", 2, 0, testData)
trigger.model.initiate()
waitcomplete()
printbuffer(1, 2, testData.sourcestatuses)
Create a reading buffer named
testData, configure the instrument to
make two measurements, and store the
readings in the buffer.
Turn on the source output.
Print the source status for the readings
in testData.
Output:
136, 136
Indicating that the status is
buffer.STAT_READBACK and
.
Also see
buffer.make() (on page 8-14)
bufferVar.clear() (on page 8-19)
buffer.delete() (on page 8-11)
print() (on page 8-92)
printbuffer() (on page 8-93)
Reading buffers (on page 3-2)
Remote buffer operation (on page 3-23)
bufferVar.sourceunits
This attribute contains the units of measure of the source.
Type TSP-Link accessible Affected by Where saved Default value
Instrument reset
Usage
readingUnits = bufferVar.sourceunits[N]
The units of measure of the source:
• Volt DC
•
The name of the reading buffer, which may be a default buffer (defbuffer1 or
) or a user-defined buffer
The reading number N; can be any value from 1 to the number of readings in the
buffer; use the bufferVar.n command to determine the number of readings in the