smu.digitize.limit[Y].autoclear
This attribute indicates if the test result for limit Y should be cleared automatically or not.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
smu.ON
Usage
value = smu.digitize.limit[Y].autoclear
smu.digitize.limit[Y].autoclear = value
• Disable: smu.OFF
• Enable:
Details
When auto clear is set to on for a measure function, limit conditions are cleared automatically after
each measurement. If you are making a series of measurements, the instrument shows the limit test
result of the last measurement for the pass or fail indication for the limit.
If you want to know if any of a series of measurements failed the limit, set the auto clear setting to off.
When this set to off, a failed indication is not cleared automatically. It remains set until it is cleared
with the clear command.
The auto clear setting affects both the high and low limits.
Example
smu.digitize.func = smu.FUNC_DIGITIZE_VOLTAGE
smu.digitize.limit[1].autoclear = smu.OFF
Turns off autoclear for limit 1 when
measuring digitize voltage.
Also see
smu.digitize.limit[Y].enable (on page 8-111)
smu.digitize.limit[Y].clear()
This attribute clears the results of the limit test defined by Y.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
smu.digitize.limit[Y].clear()
Limit number: 1 or 2