8: TSP command reference Model 2461 Interactive SourceMeter® Instrument
8-188 2461-901-01 A/November 2015
smu.source.output
This attribute enables or disables the source output.
Type TSP-Link accessible Affected by Where saved Default value
Instrument reset
Power cycle
Source configuration list
Function change
Source configuration list
Usage
sourceOutput = smu.source.output
smu.source.output = sourceOutput
Switch the source output off: smu.OFF
Switch the source output on:
Details
When the output is switched on, the instrument sources either voltage or current, as set by
smu.source.func.
Example
smu.source.output = smu.ON
Switch the source output of the instrument t
Also see
Turn the Model 2461 output on or off
smu.source.func (on page 8-184)
smu.source.offmode (on page 8-186)
smu.source.protect.level
This attribute sets the overvoltage protection setting of the source output.
Type TSP-Link accessible Affected by Where saved Default value
Instrument reset
Power cycle
Source configuration list
Source configuration list
Usage
limit = smu.source.protect.level
smu.source.protect.level = limit
The overvoltage protection value; set as smu.PROTECT_x, where x is 2V, 5V, 10V,
20V, 40V, 60V, 80V, or NONE