8: TSP command reference Model 2461 Interactive SourceMeter® Instrument
Reference Manual
8-138 2461-901-01 A/November 2015
smu.measure.configlist.query()
This function returns a list of TSP commands and parameter settings that are stored in the specified configuration
index.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
smu.measure.configlist.query(listName, index)
smu.measure.configlist.query(listName, index, fieldSeparator)
listName
A string that represents the name of a measure configuration list
index
A number that defines a specific configuration index in the configuration list
fieldSeparator
String that represents the separator for the data; use one of the following:
Comma (default): ,
Semicolon: ;
New line: \n
Details
This command recalls data for one configuration index from the specified measure configuration list
and from the source configuration list (if specified).
For additional information about the information this command recalls when using a configuration list
query command, see Instrument settings stored in a measure configuration list (on page 3-33).

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
BrandKeithley
Model2461
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals