6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument
6-74 2461-901-01 A/November 2015
[:SENSe[1]]:CONFiguration:LIST:CREate
This command creates an empty measure configuration list.
Type Affected by Where saved Default value
Command only
Recall settings
Instrument reset
Save settings Not applicable
Usage
[:SENSe[1]]:CONFiguration:LIST:CREate "<name>"
A string that represents the name of a measure configuration list
Details
This command creates an empty configuration list. To add configuration indexes to this list, you need
to use the store command.
Configuration lists are not saved when the instrument is turned off. To save a configuration list, use a
saved setup to store the instrument settings, which include defined configuration lists.
Example
:SENS:CONF:LIST:CRE "MyMeasList"
Creates a measure configuration list named
.
Also see
*SAV (on page 6-15)
Configuration lists (on page 3-30)
[:SENSe[1]]:CONFiguration:LIST:STORe (on page 6-78)
[:SENSe[1]]:CONFiguration:LIST:DELete
This command deletes a measure configuration list.
Type Affected by Where saved Default value
Usage
[:SENSe[1]]:CONFiguration:LIST:DELete "<name>"
[:SENSe[1]]:CONFiguration:LIST:DELete "<name>", <index>
A string that represents the name of a measure configuration list
A number that defines a specific configuration index in the configuration list