Model 2461
Interactive SourceMeter® Instrument Reference Manual Section 6:
SCPI command reference
2461-901-01 A/November 2015 6-79
[:SENSe[1]]:COUNt
This command sets the number of measurements to make when a measurement is requested.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings 1
Usage
[:SENSe[1]]:COUNt <n>
[:SENSe[1]]:COUNt DEFault
[:SENSe[1]]:COUNt MINimum
[:SENSe[1]]:COUNt MAXimum
[:SENSe[1]]:COUNt?
[:SENSe[1]]:COUNt? DEFault
[:SENSe[1]]:COUNt? MINimum
[:SENSe[1]]:COUNt? MAXimum
<n>
The number of measurements (1 to 300,000)
Details
This command sets the number of measurements that are made when a measurement is requested.
This command does not affect the trigger model.
This command sets the count for all measure functions.
If you set the count to a value that is larger than the capacity of the reading buffer and the buffer fill
mode is set to continuous, the buffer wraps until the number of readings specified have occurred. The
earliest readings in the count are overwritten. If the buffer is set to fill once, readings stop when the
buffer is filled, even if the count is not complete.
To get better performance from the instrument, use the Simple Loop trigger model template instead
of using the count command.

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
BrandKeithley
Model2461
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals