6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument
6-102 2461-901-01 A/November 2015
:SOURce[1]:LIST:<function>:APPend
This command adds values to the source list for the selected source function.
Type Affected by Where saved Default value
Instrument reset
Usage
:SOURce[1]:LIST:CURRent:APPend <list>
:SOURce[1]:LIST:VOLTage:APPend <list>
Current: −7.35 to 7.35 A
−
Details
This adds up to 100 values to the list created with :SOURce[1]:LIST:<function>. The new
values are added to the end of the existing values. You can have a total of 2500 values in a list, but
you must append them in groups of 100.
If the list does not exist, this command creates one.
Example
SENS:FUNC "CURR"
SENS:CURR:RANG:AUTO ON
SENS:CURR:RSEN OFF
SOUR:FUNC VOLT
SOUR:VOLT:RANG 20
SOUR:VOLT:ILIM 1
SOUR:LIST:VOLT 1, 5, 1, 5, 1, 5
SOUR:LIST:VOLT:APP 1, 5, 1, 5, 1, 5
SOUR:SWE:VOLT:LIST 1, 0.2
INIT
*WAI
TRAC:DATA? 1, 12, "defbuffer1", SOUR, READ
This example will create a source
configuration list (VoltCustomSweepList)
and source 1 V, 5 V six times and measure
the resulting current at each voltage point.
The duration of each voltage point is
200 ms.
Also see
:SOURce[1]:LIST:<function> (on page 6-101)