6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument
Reference Manual
6-102 2461-901-01 A/November 2015
:SOURce[1]:LIST:<function>:APPend
This command adds values to the source list for the selected source function.
Type Affected by Where saved Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
:SOURce[1]:LIST:CURRent:APPend <list>
:SOURce[1]:LIST:VOLTage:APPend <list>
<list>
Current: 7.35 to 7.35 A
Voltage:
105 V to 105 V
Details
This adds up to 100 values to the list created with :SOURce[1]:LIST:<function>. The new
values are added to the end of the existing values. You can have a total of 2500 values in a list, but
you must append them in groups of 100.
If the list does not exist, this command creates one.
Example
*RST
SENS:FUNC "CURR"
SENS:CURR:RANG:AUTO ON
SENS:CURR:RSEN OFF
SOUR:FUNC VOLT
SOUR:VOLT:RANG 20
SOUR:VOLT:ILIM 1
SOUR:LIST:VOLT 1, 5, 1, 5, 1, 5
SOUR:LIST:VOLT:APP 1, 5, 1, 5, 1, 5
SOUR:SWE:VOLT:LIST 1, 0.2
INIT
*WAI
TRAC:DATA? 1, 12, "defbuffer1", SOUR, READ
This example will create a source
configuration list (VoltCustomSweepList)
and source 1 V, 5 V six times and measure
the resulting current at each voltage point.
The duration of each voltage point is
200 ms.
Also see
:SOURce[1]:LIST:<function> (on page 6-101)

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
BrandKeithley
Model2461
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals