Commands Listed in Alphabetical Order
AUXIn:PRObe:
TIPtype
This command sets or queries the type of probe tip being used.
Group
Vertical
Syntax
AUXIn:PRObe:TIPtype {HBWStraightfle x|OTHer}
AUXIn:PRObe:TIPtype?
Arguments
HBWStraightflex lets the instrument know you are using a high bandwidth
straight-flex probe tip.
OTHer lets the instrument know you are not using a high bandwidth straight-flex
probe tip.
Examples
AUXIN:PROBE:TIPTYPE HBWStraightflex sets the auxiliary input probe tip
type to HBWStraightflex.
AUXIN:PROBE:TIPTYPE? might return :AUXIN:PROBE:TIP TYPE OTHER,
indicating the probe tip is not a HBWStraightfl ex type.
AUXIn:PRObe:UNIts? (Query Only)
This quer
y-only command returns a string describing the units of measure for the
probe attached to the AUX In input.
Group
Vertic
al
Syntax
AUXIn:PRObe:UNIts?
Examples
AUXIN:PROBE:UNITS? might return :AUXIN:PROBE :UNITS "V", indicating
that the units of measure for the attached probe are volts.
AUXIn:PROBEFunc:EXTAtten
This command sets the input-output ratio, of external attenuation or gain, between
the signal and the AUX In input.
The query form of this command returns the user-specified attenuation. Note
that this command deals with an attenuation factor, not a gain factor, unlike
CH<x>:PRObe? (This command returns a value independent of the external
attenuation). For example, if you specify a 20X attenuation factor, the commands
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-97