Commands Listed in Alphabetical Order
Arguments
ON enables the a
utoadjust function.
OFF disables the autoadjust function.
<NR1> = 0 disables the autoadjust function; any other value enables it.
Examples
MASK:AUTOADJUST ON enables the autoadjust function.
MASK:AUTOADJUST? might return :MASK:AU TOADJUST 1, indicating that the
autoadjust function is enabled.
MASK:AUTOAdjust:HDELTA
This command sets or returns how far autofit searches horizontally.
Autofit moves the waveform vertically and/or horizontally to reduce the number
of hits within a given mask. If infinite or variable persistence is enabled, thes e
movements will cl ear any persistant data. If Autofit makes frequent adjustments,
there might be very little or even no persistant data displayed.
Group
Mask
Syntax
MASK:AUTOAdjust:HDELTA {<NR3>}
MASK:AUTOAdjust:HDELTA?
Arguments
<NR3> is a floating point number that represents a percent of a division.
Examples
MASK:AUTOADJUST:HDELTA 0.6 sets the horizontal autofit search to
0.6 division.
MASK:AUTOADJUST:HDELTA? might return :MASK:AUTOAD JUST:HDELTA 2,
indicating that the horizontal autofit search is set to 2 divisions.
MASK:AUTOAdjust:VDELTA
This command sets or returns how far autofit searches vertically.
Autofit moves the waveform vertically and/or horizontally to reduce the number
of hits within a given mask. If infinite or variable persistence is enabled, thes e
movements will cl ear any persistant data. If Autofit makes frequent adjustments,
there might be very little or even no persistant data displayed.
Group
Mask
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-287