Commands Listed in Alphabetical Order
TRIGGER:A:PUL
SE:TIMEOUT:QUALIFY?
might return
:TRIGGER:A:PULSE:TIMEOUT:Q UALIFY OCCURS, indicating that a
trigger will occur if the instrument detects any of the timeout qualifications.
TRIGger:{A|
B}:PULse:TIMEOut:TIMe
This command sets or queries the pulse timeout trigger time (measured in
seconds). This command is equivalent to selecting Timeout Setup from the Trig
menu and se
tting a value for Timer.
Group
Trigger
Syntax
TRIGger:{A|B}:PU
Lse:TIMEOut:TIMe <NR3>
TRIGger:{A|B}:PULse:TIMEOu t:TIMe?
Related Commands
TRIGger:{A|B}:PULse:TIMEOut?
Arguments
<NR3> argument specifies the timeout period in seconds.
Examples
TRIGGER:A:PULSE:TIMEOUT:TI ME 3.134E-6 sets the timeout to 3.134 µs.
TRIGGER:A:;PULSE:TIMEOUT:T IME? might return
:TRIGGER:A:PULSE:TIMEOUT:T IME 2.0000E-9, indicating that the timeout
is currently set to 2 ns.
TRIGger:{A|B}:PULse:TRANsition? (Query Only)
This query-only command returns delta time, polarity, and both upper and lower
threshold limits for the transition time trigger. This command is equivalent to
selecting Transition Setup from the Trig menu and then viewing the current
transition settings.
Group
Trigger
Syntax
TRIGger:{A|B}:PULse:TRANsi tion?
Related Commands
TRIGger:A:TYPe, TRIGger:{A|B}:PULse:CLAss, TRIGger:{A|B}:PULse:
GLItch?, TRIGger:{A|B}:PULse:RUNT?, TRIGger:{A|B}:PULse:TIMEOut?,
TRIGger:{A|B}:PULse:WIDth?
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-535