Commands Listed in Alphabetical Order
Arguments
<NR3> is the hig
h reference level, ranging from 0 to 100%. The default high
reference level is 90%.
Examples
MEASUREMENT
:MEAS1:REFLEVEL:PERCENT:HIG H 95
sets the high reference
level to 95% of HIGH.
MEASUREMENT:MEAS1:REFLEVE L:PERCENT:HIGH? might return
:MEASUREMENT:MEAS1:REFLEV EL:PERCENT:HIGH 90, indicating that the
percentage high reference level is set to 90% of HIGH.
MEASUrement:MEAS<x>:REFLevel:PERCent:LOW
This command sets or queries the percentage (where 100% is equal to HIGH) used
to calculate the low reference level when MEASUrement:MEAS<x>:REFLevel:
METHod is set to Percent. This command is equivalent to selecting Reference
Levels from the Measure menu, and then e ntering the Percentage Low Ref value.
Measurements are specified by x, which ranges from 1 through 8.
NOTE. This command affects the results of rise and fall measurements.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:REFLe vel:PERCent:LOW <NR3>
MEASUrement:MEAS<x>:REFLe vel:PERCent:LOW?
Related Commands
MEASUrement:MEAS<x>:REFLevel:METHod
Arguments
<NR3> is the low reference level, ranging from 0 to 100%. The default low
reference level is 10%.
Examples
MEASUREMENT:MEAS1:REFLEVE L:PERCENT:LOW 15 sets the high reference
level to 15% of HIGH.
MEASUREMENT:MEAS1:REFLEVE L:PERCENT:LOW? might return
:MEASUREMENT:MEAS1
:REFLEVEL:PERCENT:LOW 10
, indicating that the
percentage high reference level is set to 10% of HIGH.
MEASUrement:MEAS<x>:REFLevel:PERCent:MID<x>
This command sets or queries the percentage (where 100% is equal to HIGH)
used to calculate the mid reference level for the second waveform specified
2-392 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual