Commands Listed in Alphabetical Order
TRIGger:A:CO
MMunication:STANdard
This command sets or queries the standard that identifies the code and bit rate.
The bit rate is used to compute the Unit Interval, which is the inverse of the bit
rate. The Uni
t Interval influences time skew in an Eye Diagram, where you
perform post processing on AMI isolated pulses, and pulse width settings if CMI.
Conditions
This command is valid only if the instrument has Option MTM is installed.
Group
Trigger
Syntax
TRIGger:A:COMMunication:ST ANdard
{ATAG1|ATAG2|ATAG3|CLOCKCo ax|
CLOCKSymmetrical|Custom|D1 |D2|DS0Contra|
DS0Double| DS0Single|DS0Timing|DS1|DS1A|
DS1C|DS2| DS2RATECoax|DS2RATESymmetrical|
DS3|DS4NA|E1|E2|E3|E4|ENET 100|ENET1250|
ENETXAUI| FC133|FC266|FC531|FC1063|FC2125 |FC4250|
FST|FW1394BS400B|FW1394BS1 600B|HST|INF2_5G|
OC1|OC3| OC12|OC48|OC48_FEC|PCIEXPRESS|RA TE32Mbit|
RATE97Mbit|RIO_500M|RIO_75 0M|
RIO_1G|RIO_2G|RIO_1_5G|RIO _SERIAL_1G|
RIO_SERIAL_2G|RIO_SERIAL_3 G|SAS1_5|SAS3_?|
SFI5_2|SFI5_3|STM0_CMI|STM 0_HDBX|STM1E|STS1| STS3|
TFI5_2|TFI5_3|VIDEO270|VID EO292M|VIDEO360|VSROC192}
TRIGger:A:COMMunication:ST ANdard?
Related Commands
MASK:STANdard
Arguments
The following table shows various standards and their corresponding parameters.
Table 2-19: Communication Trigger Standards
Argument Code Bit Rate Mask Type Models
ATAG1
NRZ
1.5 Gb/s Serial ATA ≥2.5 GHz
ATAG2
NRZ
3.0 Gb/s Serial ATA
ATAG3
NRZ
6.0 Gb/s Serial ATA
CLOCK Coax
None
2.048 Mb/s ITU-T G.703
All
CLOCK Symmetrical
None
2.048 Mb/s ITU-T G.703
All
Custom AMI, B3ZS,
B6ZS,
B8ZS, CMI,
HDB3,
MLT3, NRZ
1.544 Mb/s
All
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-563