Commands Listed in Alphabetical Order
Examples
CURSOR:WAVEFO
RM:STYLE L INES
sets the cursor style to lines.
CURSOR:WAVEFORM:STYLE? might return :CURSOR:WAV EFORM:STYLE X,
indicating th
at the style for waveform cursors is X.
CURSor:WAVEform:UNIts
This command sets or queries the units for waveform cursors.
Group
Cursor
Syntax
CURSor:WAVEform:UNIts BASe
CURSor:WAVEform:UNIts?
Arguments
BASe sets cursor units to base.
Examples
CURSOR:WAVEFORM:UNITS BASe sets the waveform cursor units to base.
CURSOR:WAVEFORM:UNITS? might return :CURSOR:WAVEFORM:UNITS BASE,
indicating that the units for waveform cursors are base.
CURSor:WAVEform:VDELTA? (Query Only)
This query-only command returns the vertical difference between the waveform
cursors. This is the absolute value of the vertical position of the first cursor minus
the vertical position of the second cursor. This is equivalent to the waveform
delta-voltage readout value.
Group
Cursor
Syntax
CURSor:WAVEform:VDELTA?
Examples
CURSOR:WAVEFORM:VDELTA? might return :CURSOR:WAVEFORM:VDELTA
1.06E-3
, indicating t
hat the voltage between the waveform cursors is 1.06 mV.
CURSor:XY? (Query Only)
This query-only command returns all of the XY cursor p arameters.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-143