Commands Listed in Alphabetical Order
Transition Set
up from the Trig menu and choosing the Trigger When Transition
Time setting.
Group
Trigger
Syntax
TRIGger:{A|B}:PULse:TRANsi tion:WHEn {FASTERthan|SLO WERthan}
TRIGger:{A|B}:PULse:TRANsi tion:WHEn?
Related Commands
TRIGger:{A|B}:PULse:TRANsition:DELTATime, TRIGger:{A|B}:PULse:
TRANsition:POLarity, TRIGger:{A|B}:PULse:TRANsition:THReshold?
Arguments
FASTERthan sets the trigger to occur when the transitioning signal is faster than
the set volts/second rate.
SLOWERthan sets the trigger to occur when the transitioning signal is slower
than the set volts/second rate.
Examples
TRIGGER:A:PULSE:TRANSITION :WHEN SLOWERTHAN sets the trigger to occur
when the transitioning signal is slower than the set volts/second rate.
TRIGGER:A:PULSE:TRANSITION :WHEN? might return
:TRIGGER:A:PULSE:TRANSITIO N:WHEN FASTERTHAN, indicating
that the transition triggers when the transitioning signal is faster than the set
volts/second rate.
TRIGger:{A|B}:PULse:WIDth? (Query Only)
This query-only command returns the width parameters for the pulse width
trigger. This command is equivalent to selecting Width Setup from the Trig menu
and then viewing the current pulse width trigger Lower Limit, Upper Limit, Trig
When and Polarity settings.
Group
Trigger
Syntax
TRIGger:{A|B}:PULse:WIDth?
Related Commands
TRIGger:{A|B}:PULse:WIDth:HIGHLimit, TRIGger:{A|B}:PULse:WIDth:
LOWLimit, TRIGger:{A|B}:PULse:WIDth:POLarity, TRIGger:{A|B}:PULse:
WIDth:WHEn
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-541