Commands Listed in Alphabetical Order
Syntax
WFMOutpre:XIN
cr?
Related Commands
DATa:SOUrce, WFMOutpre:XUNit?
Examples
WFMOUTPRE:XINCR? might return :WFMOUTPRE:XINCR 10.0000E-6,
indicating that the horizontal sampling interval is 10 µs/point (500 µ s/div).
WFMOutpre:XUNit? (Query Only)
This query-only command returns the horizontal units for the waveform specied
by the DATa:SOUrce command.
An error is reported if the DATa:SOUrce waveform does not exist.
Group
Waveform Transfer
Syntax
WFMOutpre:XUNit?
Related Commands
DATa:SOUrce
Examples
WFMOUTPRE:XUNIT? might return :WFMOUTPRE:XUNIT "HZ", indicating that
the horizontal units for the waveform are in Hertz.
WFMOutpre:XZEro? (Query Only)
This query-only command returns the sub-sample time between the trigger
sample (designated by PT_OFF) and the occurrence of the actual trigger for
the waveform specied by the DATa:SOUrce command. This value is in units
of WFMOutpre:XUNit.
An error is reported if the DATa:SOUrce waveform does not exist.
NOTE. During steady state operation, when all control changes have settled and
triggers are arriving on a regular basis, this is the only part of the preamble that
changes on each acquisition.
Group
Waveform Transfer
Syntax
WFMOutpre:XZEro?
2-646 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual

Table of Contents

Other manuals for Tektronix DPO70000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000 Series and is the answer not in the manual?

Tektronix DPO70000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals