Commands Listed in Alphabetical Order
Arguments
ON enables the r
emote display of waveform and other graticule data.
OFF disables the remote display of waveform and other graticule data.
NR1 = 0 disables remote display of waveform and other graticule data; any other
value enables remote display of waveform and other graticule data.
Examples
DISPLAY:SHOWREMOTE ON enables remote display of waveform and o ther
graticule data.
DISPLAY:SHOWREMOTE? might return :DISPLAY:SH OWREMOTE 1, indicating
that the remote display of waveform and other graticule data is currently enabled.
DISplay:STYle
This command sets or queries how the data is displayed for normal and Fas tAcq
modes. This command is equivalent to selecting Display Style from the Display
menu and choosing a style.
Group
Display Control
Syntax
DISplay:STYle {VECtors|DOTs|INTENSIFied}
DISplay:STYle?
Arguments
VECtors connects adjacent data points. New points immediately replace old ones.
DOTs displays individual data points. New points immediately replace old ones.
INTENS
IFied
causes the display to show interpolated samples with dark spaces
(Only the “real” samples are displayed).
When F
astAcq mode is enabled, intensified samples display as dots. However,
turning off FastAcq mode causes the display style to snap back to its previous
setting.
Examples
DISPLAY:STYLE VECT ORS sets the display to connect adjacent d ata points.
DISPLAY:STYLE ? might return :DISPLAY:STYLE INTENSIFIED, indicating
that interpolated samples are not displayed.
DISplay:TRIGBar
This command controls or queries the display of the trigger-level indicator bars.
Indicator bars show where the trigger voltage level is set.
2-198 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual