Commands Listed in Alphabetical Order
MEASUrement:
IMMed:REFLevel:ABSolute:HIGH
This command sets or queries the high reference level, and is the upper reference
level when MEASUrement:IMMed:REFLevel:METHod is set to Absolute. Note
that this com
mand affects the results of rise and fall measurements.
NOTE. This command affects the results of period, frequency, delay, and all
cyclic measurements.
Group
Measurement
Syntax
MEASUrement:IMMed:REFLeve l:ABSolute:HIGH <NR3>
MEASUrement:IMMed:REFLeve l:ABSolute:HIGH?
Related Commands
MEASUrement:IMMed:REFLevel:METHod
Arguments
<NR3> is the high reference level, in volts. The default is 0.0 V.
Examples
MEASUREMENT:IMMED:REFLEVE L:ABSOLUTE:HIGH 1.71 sets the high
reference level to 1.71 V.
MEASUREMENT:IMMED:REFLEVE L:ABSOLUTE:HIGH? might return
:MEASUREMENT:IMMED:REFLEV EL:ABSOLUTE:HIGH 1.7100E+00, indicating
that the a bsolute high reference level is set to 1.71 V.
MEASUrement:IMMed:REFLevel:ABSolute:LOW
This command sets or querie s the low reference level, and is the zero percent level
when MEASUrement:IMMed:REFLevel:METHod is set t o Absolute. Note that
this command affects the results of rise and fall measurements.
NOTE. This command affects the results of period, frequency, delay, and all
cyclic measurements.
Group
Measurement
Syntax
MEASUrement:IMMed:REFLeve l:ABSolute:LOW <NR3>
MEASUrement:IMMed:REFLeve l:ABSolute:LOW?
2-372 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual