Commands Listed in Alphabetical Order
TRIGger:{A|B
}:LEVel
This command sets or queries the level for the trigger. This command is
equivalent to selecting H oldoff from the Trig menu and then viewing or setting the
trigger Leve
l or s electing B Event (Delayed) Trigger Setup from the Trig menu
and setting the B Trig Level voltage.
Group
Trigger
Syntax
TRIGger:{A|B}:LEVel {ECL|TTL|<NR3>}
TRIGger:{A|B}:LEVel?
Related Comm
ands
TRIGger:{A|B}:LEVel, TRIGger:B, TRIGger:{A|B}:EDGE:SOUrce
TRIGger:B:STATE
Arguments
ECL specifies the ECL high level.
TTL specifies the TTL high level.
<NR3> specifies the trigger level in user units (usually vo
lts).
Examples
TRIGGER:A:LEVEL TTL sets the A edge trigger to TTL high level.
TRIGGER:A:LEVEL? might return :TRIGGER:A:LEVel 1.3000E+00,
indicating that the A edge trigger is set to 1.3 V.
TRIGger:{A|B}:LEVel:CH<x>
This comma nd sets or queries the CH<x> trigger level for
TRIGGER:LVLSRCPREFERENCE SRCDEPENDENT mode.
The CH<x> range is 1 to 4.
Group
Trigger
Syntax
TRIGger:{A|B}:LEVel:CH<x> {ECL|TTL|<NR3>}
TRIGger:{A|B}:LEVel:CH<x> ?
Arguments
ECL specifies the ECL high level.
TTL specifies the TTL high level.
<NR3> specifies the trigger level in user units (usually volts).
2-502 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual