Commands Listed in Alphabetical Order
NONe takes the i
nstrument off the MultiScope trigger network.
Examples
TRIGGER:MULTISCOPE:ROLE Producer sets the instrument to supply triggers
to the MultiScope trigger n etwork.
TRIGGER:MULTISCOPE:ROLE? might return :TRIGGER:MULTISCOPE: ROLE
CONSUMER
, indicating that the instrument consumes triggers.
TRIGger:SENSITivity
This command set or queries the state of trigger sensitivity. When on, the
instrument calculates a rolling average for trigger position enhancement to
suppress trigger jitter in noisy signals. This is equivalent to selecting Higher
Sensitivity for N oisy Signals from the Trigger Mode menu. This is available only
when TRIGger:ENHanced or the equivalent check box control is also on.
Group
Trigger
Syntax
TRIGger:SENSITivity {<NR1>|OFF|ON}
TRIGger:SENSITivity?
Related Commands
TRIGger:ENHanced
Arguments
<NR1> = 0 disables rolling averaging, any other value enables rolling averaging
for trigger position enhancement.
OFF disables rolling averaging for trigger position enhancement.
ON enables rolling averaging for trigger position enhancement.
Examples
TRIGGER:SENSITIVITY ON enables the trigger sensitivity feature when
TRIGger:ENHanced is enabled.
TRIGGER:SENSITIVITY? m ight return:TRIGGER:SENSITIVITY 1, indicating
the
TRIGGER:SENSITIVITY feature i s enabled.
TRIGger:STATE? (Query Only)
This query-only command returns the current state of the triggering system. This
command is equivalent to viewing the trigger status LEDs on the instrument
front panel.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-623