Commands Listed in Alphabetical Order
Examples
MASK:TEST:SRQ
:COMPLETION ON
turns on pass/fail SRQ on completion.
MASK:TEST:SRQ:COMPLETION? might return :MASK:TEST:SRQ:COMPLETION
1
, indicating
that the pass/fail SRQ on completion is turned on.
MASK:TESt:SRQ:FAILure
This command sets or returns the SRQ status on pass/fail test failure mode. When
enabled, this command cause s the instrument to send a GPIB SRQ command
when the pass/fail status changes to ‘Failing’. In other words, it sends an SRQ
when a fail
ure occurs during the mask pass/fail test.
Group
Mask
Syntax
MASK:TESt:SRQ:FAILure {OFF|ON|<NR1>}
MASK:TESt:SRQ:FAILure?
Arguments
ON turns on the pass/fail SRQ on failure.
OFF turns off the pass/fail SRQ on failure. This is the default.
<NR1> =0turnsoffthepass/failSRQonfailure, and any other integer turns on
the pass/fail SRQ on failure.
Examples
MASK:TEST:SRQ:FAILURE OFF turns off the pass/fail SRQ on failure.
MASK:TEST:SRQ:FAILURE? might return :MASK:TEST:SRQ :FAILURE 1,
indicating that the pass/fail SRQ on failure is turned on.
MASK:TESt:STATE
This command sets or returns the state of mask pass/fail testing. When enabled,
this command resets the pass/fail status information a nd starts pass/fail mask
testing. This command also generates an ‘Operation Complete’ message when
pass/fail testing completes. Refer to BUSY? and *OPC commands for information
on synchronizing a controller w ith pass/fail testing.
Group
Mask
Syntax
MASK:TESt:STATE {OFF|ON|<NR1>}
MASK:TESt:STATE?
2-332 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual