Commands Listed in Alphabetical Order
Syntax
TRIGger:{A|B}
:PULse:TRANsition:THReshold: HIGH <NR3>
TRIGger:{A|B}:PULse:TRANs ition:THReshold:HIGH?
Related Commands
TRIGger:{A|B}:PULse:TRANsition:THReshold:LOW
Arguments
<NR3> specifies the upper threshold, in volts.
Examples
TRIGGER:A:PULSE:TRANSITIO N:THRESHOLD:HIGH 120E-3 sets the upper
limit of the pulse transition trigger to 120 mV.
TRIGGER:A:PULSE:TRANSITIO N:THRESHOLD:HIGH? might return
:TRIGGER:A:PULSE:TRANSITI ON:THRESHOLD:HIGH 2.0000E +00,
indicating that the upper limit of the p ulse transition trigger is 2 V.
TRIGger:{A|B}:PULse:TRANsition:THReshold:LOW
This command sets o r queries the lower (most negative) transition trigger
threshold. This command is equivalent to selecting Transition Setup from the Trig
menu and setting the desired Lower Level voltage.
Group
Trigger
Syntax
TRIGger:{A|B}:PULse:TRANs ition:THReshold:LOW <NR3>
TRIGger:{A|B}:PULse:TRANs ition:THReshold:LOW?
Related Commands
TRIGger:{A|B}:PULse:TRANsition:THReshold:HIGH
Arguments
<NR3> specifies the lower threshold, in volts.
Examples
TRIGGER:A:PULSE:TRANSITIO N:THRESHOLD:LOW 20E-3 sets the lower limit
of the transition trigger to 20 mV.
TRIGGER:A:PULSE:TRANSITIO N:THRESHOLD:LOW? might return
:TRIGGER:A:PULSE:TRANSITI ON:THRESHOLD:LOW 50.0000E-03,
indicating that the lower limit of the transition trigger is 50 mV.
TRIGger:{A|B}:PULse:TRANsition:WHEn
This command sets or queries whether to check for a transitioning signal that
is faster or slower than the specified delta time. This is equivalent to selecting
2-540 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual