Commands Listed in Alphabetical Order
Arguments
OCCurs specifie
s a trigger if any detectable event occurs.
LOGIc specifies a trigger if the individual channel qualifications meet the
logic pattern
s and thresholds set by the TRIGger:{A|B}:LOGIc:INPut:CH<x>
commands.
Examples
TRIGGER:A:LOGIC:SETHOLD:Q UALIFY OCCURS sets the trigger qualifier to
OCCurs, which initiates a trigger i f the instrument detects any setup and hold
trigger qualifications.
TRIGGER:A:LOGIC:SETHOLD:Q UALIFY? might return
:TRIGGER:A:LOGIC:SETHOLD: QUALIFY OCCURS, indicating that a
trigger will occur if the instrument detects any of the setup and hold qualifications.
TRIGger:{A|B}:LOGIc:SETHold:SETTime
This command sets or queries the setup time for setup and hold violation
triggering. This command is equivalent to selecting Setup/Hold Setup from the
Trig menu and then setting the desired Setup Time.
Group
Trigger
Syntax
TRIGger:{A|B}:LOGIc:SETHo ld:SETTime <NR3>
TRIGger:{A|B}:LOGIc:SETHo ld:SETTime?
Related Commands
TRIGger:{A|B}:LOGIc:SETHold:HOLDTime
Arguments
<NR3> specifies the setup time for setup and hold violation triggering.
Examples
TRIGGER:A:LOGIC:SETHOLD:S ETTIME 3.0E-6 specifies that the current setup
time for setup and hold trigger is 3 µs.
TRIGGER:A:LOGIC:SETHOLD:S ETTIME? might return
:TRIGGER:A:LOGIC:SETHOLD: SETTIME 2.0000E-09, indicating
that the current setup time for setup and hold trigger is 2 ns.
TRIGger:{A|B}:LOGIc:STATE? (Query Only)
This query-only command returns the data input and trigger criteria for the logic
trigger. This command is e qu
ivalent to selecting Logic State from the Trig menu
and then viewing the current logic state settings.
2-516 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual