Commands Listed in Alphabetical Order
Arguments
EDGE is a normal
trigger. A trigger event occurs when a signal passes through
a specified voltage level in a specified direction and is controlled by the
TRIGger:A:
EDGE commands.
LOGIc specifi es that a t rigger occurs when specified conditions are met and is
controlled by the TRIGger:A:
LOGIc commands.
PULse specifies that a trigger occurs when a specifi ed pulse is found and is
controlled by the TRIGger:A:
PULse commands.
VIDeo specifies that the trigger occurs when a video signal is found. Requires an
instrument with video hardware.
I2C specifies that a trigger occurs when an Inter-IC Control signal is found.
CAN speci
fies that a trigger occurs when a Controller Area Network frame signal
is found.
SPI spec
ifies that a trigger occurs when a Serial Peripheral Interface signal is
found.
COMMun
ication
(Option MTM) specifies that a trigger occurs when a
communications signal is found. Supports AMI, HDB3, B3ZS, B6ZS,
B8ZS, CMI, MLT3, Manchester, and NRZ encoded communications signals.
COMMunication is available only if Option MTM is installed.
SERIAL spec ifies that a trigger occurs when NRZ-encoded data is found,
providing a 32-bit serial word. This argument is available with instruments with
Option PTM.
RS232 takes a signal on a d ata source and allows you to t rigger on data within
the RS232 bitstream. The data is only one byte wide.
Examples
TRIGGER:A:TYPE EDGE sets the A trigger type to EDGE.
TRIGGER:A:TYPE? might return :TRIGGE R:A:TYPE PULSE indicating that the
A trigger type is a pulse trigger.
TRIGger:A:VIDeo? (Query Only)
DPO7000 Series only: This query-only command returns the A trigger video
parameters.
Group
Trigger
Syntax
TRIGger:A:VIDeo?
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-599