Commands Listed in Alphabetical Order
Examples
DISPLAY:TRIGT
OFF
disables the trigger T.
DISPLAY:TRIGT ? might return :TRIGT 1 indicating that the trigger T is
displayed.
DISplay:VARpersist
This command sets or queries persistence decay time, which is the approximate
decay time for a freshly struck persistence sample. This command is equivalent to
selecting Display Setup from the Display menu, selecting the Appearance tab, and
then entering the desired Persist Time.
Group
Display Control
Syntax
DISplay:VARpersist <NR3>
DISplay:VARpersist?
Related Commands
DISplay:PERSistence
Arguments
<NR3> indicates the persistence decay time setting in numeric seconds. The
persistence time ranges from 50 ms to 100 s.
Examples
DISPLAY:VARPERSIST 5E-1 sets the persistence decay time to 500 ms.
DISPLAY:VARPERSIST? might return :DISPLA Y:VARPERSIST 5.0000E-01,
indicating that persistence decay time is currently set to 500 ms.
DISplay:WAVEform
This command controls or queries the display of acquisition and reference
waveforms.
Group
Display Control
Syntax
DISplay:WAVEform {<NR1>|OFF|ON}
DISplay:WAVEform?
Arguments
OFF removes the waveforms from the display.
ON displays waveforms on the display.