Commands Listed in Alphabetical Order
NOTE. Sending I
NFInite as an argument sets 2e9 envelopes. The query response
will be 2e9 not INFInite.
Group
Acquisition
Syntax
ACQuire:NUMEnv {<NR1>|INFInite}
ACQuire:NUMEnv?
Arguments
NR1 is the number o f waveform acquisitions to be enveloped.
INFInite
Examples
ACQUIRE:NUMENV 10 species that an enveloped waveform will show the result
of combining 10 separately acquired waveforms.
ACQUIRE:NUMENV? might return :ACQUIRE:NUMENV 75, indicating that 75
acquisitions are acquired for enveloped waveforms.
ACQuire:NUMFRAMESACQuired? (Query Only)
This query-only command returns the number of frames acquired when in
FastFrame Single Sequence and acquisitions are running. When the FastFrame
single sequence stop condition is set to Stop Manually, this number counts past
the number of frames which were requested as new frames a re acquired, until
STOP is pushed. When not in FastFrame Single Sequence, this value is the same
number reported by ACQ:NUMACQ.
Group
Acquisition
Syntax
ACQuire:NUMFRAMESACQuired?
Related Commands
ACQuire:NUMACq?
Examples
ACQUIRE:NUMFRAMESACQUIRED? might return
:ACQUIRE:NUMFRAMESACQUIRED 350, indicating that 350 frames have been
acquired.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-79

Table of Contents

Other manuals for Tektronix DPO70000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000 Series and is the answer not in the manual?

Tektronix DPO70000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals