Commands Listed in Alphabetical Order
NOTE. If the ins
trument is in the Remote With Lockout State (RWLS), the UNLock
command has no effect. For more information, see the ANSI-IEEE Std 488.1-1987
Standard Digital Interface for Programmable Instrumentation, section 2.8.3 on
RL State D escriptions.
Group
Miscellaneous
Syntax
UNLock ALL
Related Commands
LOCk
Arguments
ALL specifies that all front panel buttons and knobs are unlocked.
Examples
UNLOCK ALL unlocks all front panel buttons and knobs.
VERBose
This command sets or que ries the Verbose state that controls the length of
keywords on query responses. Keywords can be both headers and arguments.
NOTE. This command does not affect IEEE Std 488.2-1987 Common Commands
(those starting with an asterisk). However, this command does make a
corresponding change in the Response Header Enable State of the opposite
interface (physical or virtual GPIB interface). Refer to Introduction for more
information.
Group
Miscellaneous
Syntax
VERBose {OFF|ON|<NR1>}
Related Commands
HEADer, *LRN?, SET?
Arguments
ON sets the Verbose state to true, which returns full-length keywords for applicable
setting queries.
OFF sets the Verbose state to false, which returns minimum-length keywords
for applicable setting queries.
<NR1>
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-625