Commands Listed in Alphabetical Order
TRIGger:{A|B
}:PULse:CLAss
This comma nd sets or queries the type of pulse on which to trigger. This
command is equivalent to selecting the setup menu for the pulse type that you
want from the
Trig menu: Glitch Setup, Width Setup, Runt Setup, Timeout Setup,
or Transition Setup.
Group
Trigger
Syntax
TRIGger:{A|B}:PULse:CLAss {GLItch|RUNT|WI Dth|
TRANsition|TIMEOut|WINdow}
TRIGger:{A|B}:PULse:CLAss?
Related Commands
TRIGger:{A|B}:PULse:GLItch?, TRIGger:{A|B}:PULse:RUNT?,
TRIGger:{A|B}:PULse:WIDth?, TRIGger:{A|B}:PULse:TIMEOut?,
TRIGger:{A|B}:PULse:TRANsition?, TRIGger:A:TYPe
Arguments
GLItch triggers when a pulse is found that is of the specied polarity and width.
These are set with the c ommands TRIGger:{A|B}:PULse:GLItch:POLarity and
TRIGger:{A|B}:PULse:GLItch:WIDth.
RUNT triggers when a pulse crosses the rst preset voltage threshold but
does not cross the second preset threshold before recrossing the rst. The
thresholds are set with the TRIGger:{A|B}:PULse:RUNT:THReshold:LOW and
TRIGger:{A|B}:PULse:RUNT:THReshold:HIGH commands.
WIDth triggers when a pulse is found that has the specied polarity and is either
inside or outside the limits as specied by TRIGger:{A|B}:PULse:WIDth:
LOWLimit and TRIGge
r:{A|B}:PULse:WIDth:HIGHLimit. The polarity is
selected using the TRIGger:{A|B}:PULse:WIDth:POLarity command.
TRANsition trigge
rs when a pulse crosses both thresholds in the same direction
as the specied polarity and the transition time between the two threshold
crossings is grea t er or less than the specied time delta.
TIMEOut triggers when the pulse train stops in the selected state for longer than
the specied time.
WINdow triggers when a pulse is found that meets the conditions
set by the A Event window trigger type, specied by the following
commands: TRIGger:{A|B}:PULse:WINdow:THReshold:HIGH,
TRIGger:{A|B}:PULse:WINdow:THReshold:LOW, TRIGger:{A|B}:PULse:
WINdow:THReshold:BOTh,andTRIGger:{A|B}:PULse:WINdow:WIDTH
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-521

Table of Contents

Other manuals for Tektronix DPO70000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000 Series and is the answer not in the manual?

Tektronix DPO70000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals