Commands Listed in Alphabetical Order
Examples
MASK:AUTOSET:
STANDARD E 1COAX
selects the E1 Coax standard mask.
MASK:AUTOSET:STANDARD? might return :MASK:AUTOSET:S TANDARD
E1COAX
,indic
ating that the E1 Coax mask is selected as the current mask standard.
MASK:AUTOSet:TRIGger
This command specifies w he ther a mask AUTOSet affects the trigger level. Other
trigger set
tings such as type of trigger are not changed by autoset. The default is
ON.
This comm
and, like all the mask autoset commands, affects only an autoset on a
standard mask, not the general instrument autoset function.
Group
Mask
Syntax
MASK:AUTOSet:TRIGger {ON|OFF|<NR1>}
MASK:AUTOSet:TRIGger?
Arguments
ON enables the autoset:trigger function.
OFF disables the autoset:trigger function.
<NR1> = 0 disables the autoset:trigger function; any other value enables it.
Examples
MASK:AUTOSET:TRIGGER ON enables the autoset:trigger function.
MASK:AUTOSET:TRIGGER? might return
:MASK:AUTOSET:TRIGGER 1, indicating that the autoset:trigger function is
enabled.
MASK:AUTOSet:USER:ONE
This command sets or queries the level 1 value used by autoset for user masks.
Group
Mask
Syntax
MASK:AUTOSet:USER:ONE <NR3>
MASK:AUTOSet:USER:ONE?
Related Commands
MASK:AUTOSet:USER:TYPe, MASK:AUTOSet:USER:ZERo
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-297