Commands Listed in Alphabetical Order
Group
Waveform Trans
fer
Syntax
WFMInpre?
Related Commands
WFMOutpre?
Examples
WFMINPRE? might return the waveform formatting as :WFMINPRE:BIT_N R
8;BN_FMTRI;BYT_NR 1; BYT_OR MSB;ENCDG BIN;NR_PT 500;PT_FMT
Y; PT_O FF 0;XINCR 2 .0000E-6;XZERO 1.7536E-6; XUNIT "s";Y MULT
1.0000E-3;YOFF 0.0000; YZERO 0.0000;YUNIT "V"
.
WFMInpre:BIT_Nr
This command sets or returns the number of bits per binary waveform point for the
waveform, as specied by the DATa:DESTination command. This specication is
meaningful only when WFMInpre:ENCdg is set to BIN.
Group
Waveform Transfer
Syntax
WFMInpre:BIT_Nr <NR1>
WFMInpre:BIT_Nr?
Related Commands
DATa:DESTination, WFMInpre:ENCdg, WFMInpre:BYT_Nr,
WFMOutpre:BIT_Nr
Arguments
<NR1> number of bits per data point can be 8, 16 ( RI, RP) or 32 (FP).
Examples
WFMINPRE:BIT_NR 16 sets the number of bits per waveform point to 16, for
incoming RI and RP binary format data.
WFMINPRE:BIT_NR ? might return :WFMINPRE:BIT_NR 8, indicating that
incoming RI or RP binary format data uses 8 bits per waveform point.
WFMInpre:BN_Fmt
This comman
d sets or queries the format o f binary data for incoming waveforms.
Group
Waveform Transfer
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-629

Table of Contents

Other manuals for Tektronix DPO70000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000 Series and is the answer not in the manual?

Tektronix DPO70000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals