Commands Listed in Alphabetical Order
Syntax
REF<x>:VERTic
al:SCAle <NR3>
REF<x>:VERTical:SCAle?
Related Commands
CH<x>:SCAle, MATH<x>:VERTical:SCAle
Arguments
<NR3> is the gain in user units per division.
Examples
REF4:VERTICAL:SCALE 100E-03 sets the Reference 4 waveform scale to
100 mV per division.
REF4:VERTICAL:SCALE? might return :REF2:VERTICAL:SCALE
1.0000e+00
, indicating that the current vertical scale setting for Reference 2
waveform is 1 V per division.
REM (No Query Form)
This command (no query form) e mbeds a comment within GPIB programs as a
means of internally documenting the programs. The instrument ignores these
embedded comment lines.
Group
Miscellaneous
Syntax
REM <QS tring>
Related Commands
NEWpass, *PUD
Arguments
<QString> is a string that can contain a maximum of 80 characters.
Examples
REM "This is a comment" is a comment string that the instrument will ignore.
ROSc:SOUrce
This command sets or queries the selected source for the time base reference
oscillator. The reference oscillator locks to this source. Depending on the
command argument that you specify, you can use an external reference or use the
internal crystal oscillator as the time base reference.
This command is also useful for synchronizing multiple instruments.
2-422 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual