Commands Listed in Alphabetical Order
MEASUrement:
IMMed:VALue? (Query Only)
This query-only command returns the value of the measurement specified by the
MEASUrement:IMMed:TYPe command. The measurement is immediately taken
on the source
(s) specified by a MEASUrement:IMMed:SOUrce<x> command.
NOTE. A change to CH<x>:SCALe will not necessarily have taken affect it
followed by this command.
Group
Measurement
Syntax
MEASUrement:IMMed:VALue?
Related Commands
MEASUrement:IMMed:TYPe, MEASUrement:IMMed:SOUrce<x>, *ESR?,
ALLEv?
Examples
MEASUREMENT:IMMED:VALUE? might return :MEASUREMENT:IMMED:V ALUE
9.9000E+37.
If the measurement has an error or wa r ning associated with it,
then an item is added to the error queue. The error can be checked for with the
*ESR? and ALLEv? comma nds.
MEASUrement:MEAS<x>? (Query O nly)
This query-only command returns all measurement parameters for the displayed
measurement specified by x, which can range from 1 through 8. This query
command is equivalent to selecting Measurement Setup from the Measure menu
and viewing the Measurements table; then choosing the Time tab, clicking the
Delay button and viewing the Delay Edge and Search Direction settings.
Group
Measurement
Syntax
MEASUrement:MEAS<x>?
Examples
MEASUREMENT:MEAS1? might return :MEASUREMENT:MEAS1:STATE
0;TYPE UNDEFINED;UNITS "V";SOURCE1 CH1;SOURCE2
CH1;SOURCE1:SIGTYPE PULSE;:MEASUREM ENT:MEAS1:SOURCE 2:SIGTYPE
PULSE;:MEASUREMENT:MEAS1:D ELAY:EDGE1 RISE;EDGE2 RISE;
DIRECTION FORWARDS;:MEASUREMENT:MEAS1:REF LEVEL:METHOD
PERCENT;ABSOLUTE:HIGH 0.0000;LOW 0.0000;MID1 0. 0000;MID2
0.0000;:MEASUREMENT:MEAS1: REFLEVEL:PERCENT:HIGH 90. 0000;LOW
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-383