Commands Listed in Alphabetical Order
TRIGger:{A|B
}:LOGIc:SETHold:CLOCk:THReshold:CH<x>
This command sets or queries the clock voltage threshold for setup and hold
trigger. This command is equivalent to selecting Setup/Hold Setup from the Trig
menu and then
setting the desired Clock Level. The value of x can range from
1 through 4.
Group
Trigger
Syntax
TRIGger:{A|B}:LOGIc:SETHo ld:CLOCk:THReshold:CH<x>
{ECL|TTL|<NR3>}
TRIGger:{A|B}:LOGIc:SETHo ld:CLOCk:THReshold:CH<x>?
Related Commands
TRIGger:{A|B}:LOGIc:SETHold:CLOCk:EDGE, TRIGger:{A|B}:LOGIc:
SETHold:CLOCk:SOUrce
Arguments
ECL specifies a preset ECL high level.
TTL specifies a preset TTL high level.
<NR3> is the clock level, in volts.
Examples
TRIGGER:A:LOGIC:SETHOLD:C LOCK:THRESHOLD: CH3 TTL specifies the
preset TTL value as the clock threshold for setup and hold trigger for Channel 3.
TRIGGER:A:LOGIC:SETHOLD:C LOCK:THRESHOLD: CH3? might return
:TRIGGER:A:LOGIC:SETHOLD: CLOCK:THRESHOLD: CH3 1.2000E+00,
indicating that the clock threshold for setup and hold trigger is 1.2 V for Channel 3.
TRIGger:{A|B}:LOGIc:SETHold:DATa? (Query Only)
This query-only command returns the voltage threshold and data source for the
setup and hold trigger. This command is equivalent to selecting Setup/Hold Setup
from the Trig menu and then viewing the current data setu ps.
Group
Trigger
Syntax
TRIGger:{A|B}:LOGIc:SETHo ld:DATa?
Related Commands
TRIGger:{A|B}:LOGIc:SETHold:CLOCk?
2-512 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual