Commands Listed in Alphabetical Order
CALibrate:IN
TERNal (No Query Form)
This command (no query form) starts the internal signal path calibration (SPC) of
the instrument. You can use the CALibrate:INTERNal:STATus? query to return
the current s
tatus of the internal signal path calibration of the instrument.
Group
Calibration
Syntax
CALibrate:INTERNal
Related Commands
CALibrate:RESults:SPC?
Examples
CALIBRATE:INTERNAL starts the internal signal path calibration of the
instrument.
CALibrate:INTERNal:STARt (No Query Form)
This command (no query form) starts the internal signal path calibration (SPC)
of the instrument. This command is the same as the CALibrate:INTERNal
command. You can use the CALibrate:INTERNal:STATus? querytoreturnthe
current status of the internal signal path calibration of the instrument.
Group
Calibration
Syntax
CALibrate:INTERNal:STARt
Related Commands
CALibrate:RESults:SPC?
Examples
CALIBRATE:INTERNAL:START starts the internal signal path calibration of the
instrument.
CALibrate:INTERNal:STATus? (Query Only)
This query-only command returns the current status of the internal signal path
calibration.
Group
Calibration
2-104 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual

Table of Contents

Other manuals for Tektronix DPO70000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000 Series and is the answer not in the manual?

Tektronix DPO70000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals