Commands Listed in Alphabetical Order
Examples
MASK:MASKPRE:
PRESAMPBITS 4
sets the current mask number of bits before
the trigger pulse to 4.
MASK:MASKPRE
:PRESAMPBITS?
might return :MASK:MASKPRE:PRESAMPBITS
8
, indicating that the presample bit value of the displayed mask is set to 8.
MASK:MASKPRE:RECOrdlength
This command s ets or returns the nominal record length for pulse mask testing.
The query form of this command returns the record length value of the displayed
mask. The set form of this command only affects the current (displayed) mask.
Group
Mask
Syntax
MASK:MASKPRE:RECOrdlength <NR1>
MASK:MASKPRE:RECOrdlength?
Arguments
<NR1> is an integer number that sets the record length.
Examples
MASK:MASKPRE:RECORDLENGTH 500 sets the current mask record length to
500 points.
MASK:MASKPRE:RECORDLENGTH? might return
:MASK:MASKPRE:RECORDLENGTH 400, indicating that the nominal record
length for pulse mask testing is set to 400.
MASK:MASKPRE:TRIGTOSAMP
This command sets or returns the time, in seconds, from the (leading edge) trigger
position to the pulse bit sampling position. The query form of this command
returns the time value of the displayed mask. The set form of this command only
affects the current (displayed) mask.
Group
Mask
Syntax
MASK:MASKPRE:TRIGTOSAMP <NR3>
MASK:MASKPRE:TRIGTOSAMP?
Related Commands
MASK:MASKPRE:WIDth
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-311