Commands Listed in Alphabetical Order
Arguments
RISe specifies t
o trigger on the rising or positive edge o f a signal.
FALL specifies to trigger on the falling or negative edge of a signal.
Examples
TRIGGER:A:COMMUNICATION:C LOCK:POLARITY RISE sets RISE as the clock
polarity.
TRIGGER:A:COMMUNICATION:C LOCK:POLARITY? might return
:TRIGGER:A:COMMUNICATION: CLOCK:POLARITY RISE, indicating that the
polarity is to trigger on the rising edge of the clock.
TRIGger:A:COMMunication:CMI:PULSEForm
This command sets or queries the CMI pulse form.
Conditions
This command is only valid if the instrument has Option MTM is installed.
Group
Trigger
Syntax
TRIGger:A:COMMunication:C MI:PULSEForm
{PLUSOne|MINUSOne|ZERO|EY Ediagram}
TRIGger:A:COMMunication:C MI:PULSEForm?
Arguments
PLUSOne triggers on a positive mark.
MINUSOne triggers on a negative mark.
ZERO triggers on the bit representing zero.
EYEdiagram positions the trigger point off screen in order to display the eye.
Examples
TRIGGER:A:COMMUNICATION:C MI:PULSEFORM PLUSONE sets triggering to
occur on a positive mark.
TRIGGER:A:COMMUNICATION:C MI:PULSEFORM? might return
:TRIGGER:A:COMMUNICATION: CMI:PULSEFORM PLUSONE, indicating that
the p ulse form is PLUSONE.
TRIGger:A:COMMunication:CODe
This command sets or queries the signal code that the communications trigger
should expect on the incoming signal.
2-560 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual