Commands Listed in Alphabetical Order
Group
TekLink
Syntax
TRIGger:MULTiscope:ALIGN:V ALue?
Examples
TRIGGER:MULTISCOPE:ALIGN:V ALUE? might return
:TRIGGER:MULTISCOPE:ALIGN: VALUE 33.000E–6, indicating that the
round trip trigger time is 3 3 μs.
TRIGger:MULTiscope:DELay? (Query Only)
This command queries the delay time, in nanoseconds, for trigger out of the
TekLink trigger out line.
Group
TekLink
Syntax
TRIGger:MULTiscope:DELay?
Examples
TRIGGER:MULTISCOPE:DELAY? might return :TRIGG ER:MULTISCOPE:DELAY
3.5000E–9
, indicating that the set TekLink trigger delay time is 3.5 ns.
TRIGger:MULTiscope:OPTion
This command sets or queries the trigger type for the TekLink trigger. This will
cause the instrument to set the hub and all instruments participating to either
AND or OR triggering.
Group
TekLink
Syntax
TRIGger:MULTiscope:OPTion {AND|OR}
TRIGger:MULTiscope:OPTion?
Arguments
AND sets the hub and all participating instruments to AND triggering.
OR sets the hub and all participating instruments to OR triggering.
Examples
TRIGGER:MULTISCOPE:OPTION OR sets the hub and all participating instruments
are set to OR triggering.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-621

Table of Contents

Other manuals for Tektronix DPO70000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000 Series and is the answer not in the manual?

Tektronix DPO70000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals