Commands Listed in Alphabetical Order
Arguments
SRCDependent s
ets the level of all trigger sources to the value you currently
select, regardless of the last value selected.
SRCIndepende
nt
sets each trigger source to the level you are currently selecting.
Examples
TRIGger:LVLSRCPREFERENCE SRCIINDEPENDENT allows you to set the level
of each source to the value you are currently selecting.
TRIGGER:LVLSRCPREFERENCE? might return :TRIGGER:LVLSRCPREFERENCE
SRCDEPENDENT
, indicating that the level of all sources changes to the value you
currently select.
TRIGger:MULTiscope
This command sets or returns the state of MultiScope triggering, either ENABle
or DISable.
Group
TekLink
Syntax
TRIGger:MULTiscope {DISable|ENable}
TRIGger:MULTiscope?
Arguments
DISable sets the MultiScope triggering state to disabled.
ENable sets the MultiScope triggering state to enabled.
Examples
TRIGGER:MULTISCOPE ENABLE allows you to u se the MultiScope triggering.
TRIGGER:MULTISCOPE? might return :TRIGGER:MULTISCO PE DISABLE,
indicating that the MultiScope triggering is disabled.
TRIGger:MULTiscope:ALIGN (No Query Form)
This command starts the TekLink trigger alignment procedure on all oscilloscopes
which are participating in MultiScope triggering. For this command to h
ave any
effect, the instrument must be participating in MultiScope triggering. Other
instruments on the TekLink network are not affected.
Group
TekLink
Syntax
TRIGger:MULTiscope:ALIGN
2-618 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual