Commands Listed in Alphabetical Order
set form of this
command only affects the user mask, regardless of the current
(displayed) mask.
Group
Mask
Syntax
MASK:USER:HTRIGPOS <NR3>
MASK:USER:HTRIGPOS?
Related Commands
MASK:USER:HSCAle
Arguments
<NR3
>
is a oating point number in the range of 0.0 to 1.0 that sets the trigger
points as a fraction of the display width, referenced from the left edge of the
graticule.
0.0 represents the left edge.
Examples
MASK:USER:HTRIGPOS 0.6 sets the nominal trigger position of the displayed
mask to 0.6.
MASK:USER:HTRIGPOS? might return :MASK:USER:HTRIGPOS 500.0000E–3,
indicating that the nominal trigger position of the displayed mask is set to 0.5.
MASK:USER:LABel
This command s ets or queries the label assigned to a user mask. This label is
displayed in the upper left-hand corner of the graticule when a user mask is
selected.
Group
Mask
Syntax
MASK:USER:LABel <string>
MASK:USER:LABel?
Arguments
<string> is an alphanumeric character string u
sedtoannotateausermasklabel.
Examples
MASK:USER:LABEL"mikesdata" assigns “mikesdata as the user mask label.
MASK:USER:LABEL? might return :M ASK:USER:LABel "mikesdata",
indicating that the label assigned to a user mask is set to “mikesdata”.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-337

Table of Contents

Other manuals for Tektronix DPO70000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000 Series and is the answer not in the manual?

Tektronix DPO70000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals