Page #202 background image
Commands Listed in Alphabetical Order
Group
Display Contro
l
Syntax
DISplay:INTENSITy:WAVEfor m:IMAGEView <NR2>
DISplay:INTENSITy:WAVEfor m:IMAGEView?
Related Commands
DISplay:INTENSITy:WAVEform:RECORDView
Arguments
<NR2> is the waveform saturation and ranges from 10 to 100 percent.
Examples
DISPLAY:INTENSITY:WAVEFOR M:IMAGEVIEW 100 sets the image view
waveform saturation to 100 percent.
DISPLAY:INTENSITY:WAVEFOR M:IMAGEVIEW? might return
:DISPLAY:INTENSITY:WAVEFO RM:IMAGEVIEW 60.0000, indicating that the
saturation level of the image view waveforms is curre
ntly set to 60 percent.
DISplay:INTENSITy:WAVEform:RECORDView
This command sets or queries the saturation level for record view waveforms.
Group
Display Control
Syntax
DISplay:INTENSITy:WAVEfor m:RECORDView <NR2>
DISplay:INTENSITy:WAVEfor m:RECORDView?
Related Commands
DISplay:INTENSITy:WAVEform:IMAGEView
Arguments
<NR2> is the waveform saturation and ranges from 10 to 100 percent.
Examples
DISPLAY:INTENSITY:WAVEFOR M:RECORDVIEW 100 sets the record view
waveform saturation to 100 percent.
DISPLAY:INTENSITY:WAVEFOR M:RECORDVIEW? might return
:DISPLAY:INTENSITY:WAVEFO RM:RECORDVIEW 60.0000, indicating that the
saturation of the record view waveforms is currently set to 60 percent.
2-190 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual

Table of Contents

Other manuals for Tektronix DPO70000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000 Series and is the answer not in the manual?

Tektronix DPO70000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals