Commands Listed in Alphabetical Order
Examples
HORIZONTAL:MO
DE AUTO
sets the horizontal mode to auto.
HORIZONTAL:MODE? might return : HORIZONTAL:MODE MANUAL, indicating that
the horizonta
l mode is manual.
HORizontal:MODE:AUTO:LIMITrecordlen
This command sets or queries the record length limit used by the auto horizontal
mode.
Group
Horizontal
Syntax
HORizontal:MODE:AUTO:LIMIT recordlen <NR1>
HORizontal:MODE:AUTO:LIMIT recordlen?
Arguments
<NR1> is the record length limit in samples.
Examples
HORIZONTAL:MODE:AUTO:LIMIT RECORDLEN 1000 sets the record length limit
to 1000 samples.
HORIZONTAL:MODE:AUTO:LIMIT RECORDLEN? might return
:HORIZONTAL:MODEL:AUTO:LIM ITRECORDLEN 1000000, indicating that the
record length limit is set to 1,000,000 samples.
HORizontal:MODE:RECOrdlength
This command sets or queries the record length.
Group
Horizontal
Syntax
HORizontal:MODE:RECOrdleng th <NR1>
HORizontal:MODE:RECOrdleng th?
Arguments
<NR1> is the record length in samples. Manual mode lets you change the record
length, while the record length is read only for Auto and Constant mode.
Examples
HORIZONTAL:MODE:RECORDLENG TH 1000 sets the record length to
1000 samples.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-261

Table of Contents

Other manuals for Tektronix DPO70000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000 Series and is the answer not in the manual?

Tektronix DPO70000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals